The Effect of Repeated Bonding on Shear Bond Strength and Cellular Toxicity of Composite Orthodontic Adhesives
Keywords:
Orthodontic adhesive, repeated bonding, shear bond strength, rebonding, cellular toxicity, cell viabilityAbstract
Background: There is a need for composite orthodontic adhesives to be strong and safe. Multiple bonding may change the bond interface and affect mechanical properties and cell responses.
Methods: In-vitro laboratory studies were done between March 2026 and August 2026. The teeth were extracted from 60 humans and randomly divided into three groups, initial bonding, first rebonding, and second rebonding. Shear bond strength was determined using a testing machine. An extract (eluate) test was performed on L929 cells (mouse fibroblasts) using the MTT assay, and the cell viability was expressed as a percentage of the untreated control and categorized according to the Dahl index.
Results: The mean shear bond strength was 18.42 ± 2.31 MPa after the initial bonding, 15.76 ± 2.48 MPa after the first rebonding, and 13.21 ± 2.57 MPa after the second rebonding. Mean cellular viability decreased from 91.6 ± 4.2% to 84.3 ± 5.1% and 76.8 ± 5.8%, respectively. There was a negative correlation between repeated bonding and bond strength and cell viability (p < .001 for both), but cell viability remained above 70% of the control (no cytotoxicity in the initial bonding group and slight cytotoxicity after rebonding according to the Dahl index).
Conclusion: The repeated bonding affected the shear bond strength and the cell viability of the composite orthodontic adhesive, with the cytotoxic response still being in the slight range. These findings should be confirmed in further studies, and optimized rebonding protocols should be developed.
References
Seifi M, Eskandarloo F, Amdjadi P, Farmany A. Investigation of mechanical properties, remineralization, antibacterial effect, and cellular toxicity of composite orthodontic adhesive combined with silver-containing nanostructured bioactive glass. BMC Oral Health. 2024 Jun 1;24(1):650.
Nookala H, JAIN RK. Comparative Evaluation of Shear Bond Strength, Adhesive Remnant Index and Cytotoxicity of Different Orthodontic Adhesives: An In-vitro Study. Journal of Clinical & Diagnostic Research. 2024 Oct 1;18(10).
Bautista-Martinez D, Scougall-Vilchis RJ, Contreras-Bulnes R, Garcia-Contreras R. Assessment of cytotoxicity, odontoblast-like differentiation, shear bond strength, and microhardness of four orthodontic adhesive composites. Journal of Oral Science. 2024;66(4):220-5.
Al Akel NE. Evaluation of the Effect of Titanium Dioxide Nanoparticles on the Shear Bond Strength of Orthodontic Bonding Agent and Composite Adhesive (Master's thesis, Marmara Universitesi (Turkey)).
Jedliński M, Mazur M, Salerno C, Grocholewicz K, Janiszewska-Olszowska J. Impact of bonding protocols and physical-mechanical properties of composite on durability and failure rate of fixed orthodontic retainers: a systematic review of laboratory studies. BMC Oral Health. 2025 Jul 26;25(1):1255.
Tivanani MV, Mulakala V, Keerthi VS. Antibacterial properties and shear bond strength of titanium dioxide nanoparticles incorporated into an orthodontic adhesive: a systematic review. International Journal of Clinical Pediatric Dentistry. 2024 Jan;17(1):102.
Pourhajibagher M, Ghorbanzadeh R, Salehi-Vaziri A, Bahador A. In vitro assessments of antimicrobial potential and cytotoxicity activity of an orthodontic adhesive doped with nano-graphene oxide. Folia Medica. 2022 Feb 28;64(1):110-6.
de Abreu NM, de Sousa FB, dos Santos Oliveira R. Effect of nanoparticle incorporation on antimicrobial activity and shear bond strength of orthodontic adhesives: a scoping review. The Saudi Dental Journal. 2025 Oct;37(7):62.
Khojah MZ, Youssef AR, Alsahafi EN. The effect of expired orthodontic bonding material on primary human gingival fibroblasts: In vitro study. Journal of Orthodontic Science. 2025 Mar 25;14:3.
Arampatzi M, Spyratou E, Sifakakis I, Efstathopoulos EP. Clinical Roles of Nanoparticles in Orthodontic Bonding Materials. Applied Sciences. 2026 Feb 17;16(4):1996.
Pourhajibagher M, Bahador A. Physico-mechanical properties, antimicrobial activities, and anti-biofilm potencies of orthodontic adhesive containing cerium oxide nanoparticles against Streptococcus mutans. Folia Medica. 2022 Apr 30;64(2):252-9.
Alhazmi AS, Elshebiny T, Alayyash A, Asiri SN, Alharbi F. Comparative Assessment of the Shear Bond Strength of Green Synthetized Titanium Dioxide, Hydroxyapatite and Chitosan Nanoparticles Integrated in Orthodontic Adhesive and Evaluation of Antibacterial Ability, an In-vitro and Laboratory Study. The Open Dentistry Journal. 2024 Nov 15;18(1).
Barylyak A, Kisała J, Fornal P, Bobitski Y, Dziedzic A, Kus-Liśkiewicz M. Physical, mechanical and antibacterial properties of the orthodontic adhesive doped with S-TiO2. Scientific Reports. 2024 Aug 1;14(1):17862.
Adham R, Ab Rahman N, Ponnuraj KT. Assessment of DNA damage of oral mucosa due to ceramic bracket using comet assay and mutagenicity of orthodontic bonding system using Ames test. Journal of Dentistry Indonesia. 2023;30(1):33-40.
Shallal W, Zbidi ND, Hasan SM, Hussein AD, Hafith AN. Investigation of the effect of clove nanoparticles on antibacterial features and shear bond strength of orthodontic adhesive bonded to human premolars teeth. J Nat Sci Biol Med. 2023 Jul;14(2):288-98.
Ohlsson E, Bolay C, Arabulan S, Galler KM, Buchalla W, Schmalz G, Widbiller M. In-vitro-cytotoxicity of self-adhesive dental restorative materials. Dental Materials. 2024 Apr 1;40(4):739-46.
Prasad T, Pawar R, Ganiger C, Ronad Y, Phaphe S, Mane P, Patil S. The impact of orthodontic adhesive containing resveratrol, silver, and zinc oxide nanoparticles on shear bond strength: An in vitro study. Cureus. 2024 Aug 31;16(8).
Sycińska-Dziarnowska M, Szyszka-Sommerfeld L, Ziąbka M, Spagnuolo G, Woźniak K. Antimicrobial efficacy and bonding properties of orthodontic bonding systems enhanced with silver nanoparticles: a systematic review with meta-analysis. BMC Oral Health. 2024 Nov 5;24(1):1342.
International Organization for Standardization. ISO 10993-5:2009. Biological evaluation of medical devices. Part 5: Tests for in vitro cytotoxicity. Geneva: ISO; 2009.
International Organization for Standardization. ISO 10993-12:2021. Biological evaluation of medical devices. Part 12: Sample preparation and reference materials. Geneva: ISO; 2021.
Dahl JE, Frangou-Polyzois MJ, Polyzois GL. In vitro biocompatibility of denture relining materials. Gerodontology. 2006;23(1):17-22.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Authors

This work is licensed under a Creative Commons Attribution 4.0 International License.